Fix "Electric Tracer in an Unsafe Status" Error

If you're encountering the "Electric Tracer in an Unsafe Status" error on your Keyline machine, follow these specific troubleshooting steps based on your machine model:

Gymkana

  1. Check Tracer Type: Ensure the T08 tracer is properly installed in the tracer shaft.

  2. Tracer Setup Check: If performing a tracer setup, confirm the T00 tracer is installed in the cutting shaft.

Ninja Total

Cause: This error is often caused by metal chips from cutting keys accumulating in the carriage and causing a short.
Solution:

  1. Remove Carriage Cover: Take off the front black carriage cover.

  2. Clean the Carriage: Use a vacuum to remove any metal chips or debris.

  3. Persistent Issue: If the error remains, contact Keyline Tech Support for repair: techsupport@keyline-usa.com.

Pro Source

  1. Check Tracer Type: Ensure the T08 tracer is correctly installed in the tracer shaft.

  2. Tracer Setup Check: If performing a tracer setup, confirm the T00 tracer is in the cutting shaft.

Ninja Laser

Cause: This error usually happens when metal chips from key cutting collect inside the carriage, leading to an electrical short.
Solution:

  1. Remove Carriage Cover: Take off the front black carriage cover.

  2. Clean the Carriage: Use a vacuum to clear away any metal chips or particles.

  3. Persistent Issue: If the problem continues, contact Keyline Tech Support for repair: techsupport@keyline-usa.com.

Messenger

For the Messenger machine, contact Keyline Tech Support for assistance: techsupport@keyline-usa.com.

Still Need Help? Reach out to techsupport@keyline-usa.com for additional support.